AZ 86031 DO Salinity Compensation: Check P62 Before Recalibrating
A dissolved-oxygen reading that disagrees with another instrument does not automatically mean the probe needs recalibration. On the AZ 86031, first check whether the salinity used for the DO measurement matches the water being tested. A visible salinity result elsewhere on the screen is not evidence that the DO compensation setting has updated itself.
This guide addresses that specific handover problem: moving a shared meter between fresh and saline water, or comparing its result with another system, without losing track of the compensation inputs. Treat the configuration record as part of the measurement record.
Measurement and compensation are separate checks
Hamilton’s oxygen measurement guide explains that dissolved salt affects oxygen solubility and that salinity compensation can require a manually entered value from a separate measurement. The practical implication is that an oxygen concentration result needs context: the sample, temperature, compensation inputs and units must travel with the number.
Do not transfer another manufacturer’s setup method to the AZ. Vernier’s own technical note, for example, describes different compensation workflows for its different probe families. That is a useful reminder to check the exact instrument manual, rather than assuming all multi-parameter meters link their channels in the same way.
What the AZ 86031 manual actually requires
The AZ operating manual describes manual DO salinity compensation under P60. P62 accepts salinity in ppt, using the value measured by the conductivity probe; the documented range is 0.0–42.00 ppt. P63 separately sets height above sea level, from 0–3500 m in 100 m steps. These are distinct settings, not interchangeable correction factors.
The ProSense product listing also makes probe configuration explicit: measurement functions depend on the connected probes. Confirm that the supplied arrangement includes the probes needed for your method. A photograph showing several probes does not establish the contents of every order.

A controlled check before changing calibration
Use the following sequence as a recordkeeping and troubleshooting checklist alongside the manufacturer’s complete preparation, calibration and measurement instructions. It is not a substitute for the site’s approved water-testing method.
Identify the measurement. Record the meter and probe identities, sampling point, date, time and required output unit. Keep mg/L and percentage saturation clearly labelled in worksheets and comparison reports.
Preserve the original evidence. Before editing settings, record the displayed DO result, sample temperature and existing compensation entries. A photograph of the display and a written configuration record can prevent uncertainty about what was changed.
Establish the sample’s salinity. Use the approved method and an appropriate verified measurement. Keep the measured salinity and its unit with the sample record; do not substitute a site nickname such as “brackish” or a remembered value from yesterday.
Check the AZ compensation entry. Compare P62 with the current sample salinity and check P63 against the site elevation. Follow the manual to enter and confirm appropriate values. Record both the original and revised entries.
Repeat under controlled conditions. Use the same approved sampling arrangement and allow the instrument to stabilise as instructed. Record the repeat result without overwriting the original. If comparing two instruments, document both configurations and take readings close enough in time for the process comparison to be meaningful.
Decide what remains unresolved. If disagreement persists, review calibration status, probe condition and sampling conditions using the relevant manuals. Escalate unexplained differences before using the result to justify a process adjustment.
Worked example: a freshwater setting travels with the meter
Consider a hypothetical maintenance round, not product test data. A shared meter is used at a freshwater point, then taken to a second point where the approved salinity measurement is 18 ppt. The operator records that the DO compensation entry still reads 0 ppt.
The useful finding is a configuration mismatch. Record it, follow the manual to set the appropriate input, and repeat the measurement. Do not invent the corrected oxygen concentration or apply a blanket percentage adjustment. If the first result has already entered a report, retain its history and flag it for review under the site’s data-quality procedure.
Keep the correction from becoming a new error
Avoid changing calibration simply to force agreement with a reference display. Also avoid silently correcting historical readings: the earlier sample conditions and settings may not be recoverable. Mark uncertain records rather than presenting reconstructed values as measured facts.
For the next handover, add compensation inputs to the field sheet and require operators to check them when changing sampling locations. Ask the responsible technical person to define comparison tolerances and acceptance criteria for the application. A handheld spot check alone does not establish that a treatment process meets its operating or discharge requirements.