Pressure Calibration: Keep Rising and Falling Errors Separate
A pressure-calibration spreadsheet can show an acceptable average while concealing an unacceptable individual reading. If the instrument reads differently at the same reference pressure depending on whether pressure is rising or falling, combining those observations removes information needed for the decision. Keep the direction, reference reading and instrument indication together throughout the record.
What the two directions tell you
Beamex explains pressure hysteresis as dependence on pressure history: the indication at a given applied pressure can differ between an increasing and a decreasing run. Its guidance recommends testing both directions and avoiding overshoot when approaching test points. A single rising run cannot reveal the separation between the two paths.
For this review, treat hysteresis as a measurement result to investigate, rather than proof of a particular damaged component. Use the instrument manufacturer’s definition and approved calibration method when reporting it. Pressure-switch trip-to-reset differential is a different task; this article concerns gauges and transmitters that report a measured value.

The pictured LFT6100 digital pressure gauge illustrates local pressure indication. Confirm the ordered range, pressure type, connection and applicable calibration evidence for the actual unit. Its digital display alone does not establish suitability as the reference instrument for a calibration.
Build a record that cannot lose direction
Give each observation its own row. Useful columns are instrument ID, cycle number, direction, nominal test point, actual reference pressure, actual indication, calculated error and applicable limit. Retain the original values when a report or dashboard later summarises the results. Label the units and pressure type explicitly.
Calculate indicated error as instrument indication minus reference pressure, with any authorised corrections clearly identified. Compare paired observations at the same applied reference pressure using the approved method. Two rows with the same nominal target are not necessarily observations at identical pressure: preserve the actual reference values so that distinction can be checked.
Keep a direction field mandatory when importing data. For example, use separate entries such as ‘rising, cycle 1’ and ‘falling, cycle 1’, rather than replacing both with ‘8 bar’. If the instrument output is converted from current or voltage to pressure, retain the original electrical measurement and the conversion settings as well.
Worked example: the average conceals a limit exceedance
Consider a hypothetical 0–16 bar gauge assessed at a reference pressure of exactly 8.000 bar. The rising indication is 7.940 bar and the falling indication is 8.020 bar. Assume the approved worksheet uses an illustrative ±0.040 bar error limit at this point. These invented values are an arithmetic example, not a specification or test result for the photographed gauge.
The rising error is 7.940 − 8.000 = −0.060 bar. The falling error is 8.020 − 8.000 = +0.020 bar. The indication separation at this point is 0.080 bar; expressed against the hypothetical 16 bar span, that is 0.5% of span.
Averaging the indications gives 7.980 bar, or an error of −0.020 bar. That average lies inside the illustrative limit, even though the rising observation does not. A worksheet that retains only the average has hidden the relevant result. This simplified comparison excludes measurement uncertainty; an actual conformity decision must apply the agreed decision rule and uncertainty treatment.
Check the measurement conditions before interpreting the pattern
Beamex’s pressure-gauge calibration guidance highlights pressure-medium compatibility, contamination, mounting position, leaks and stabilisation. Rapid pressure changes can produce temperature-related pressure changes in gas systems. Follow the exact equipment procedures and allow the required stabilisation before recording results; do not substitute a universal waiting time.
Have competent personnel establish an isolated test arrangement with suitable pressure-rated equipment and compatible media. Follow the authorised preparation and depressurisation method. Never loosen a pressurised connection. This guide concerns records and interpretation, not instructions for working on a live process.
Mistakes to catch before approving the report
Check that an export has not merged direction rows, rounded away the separation or changed an error into an absolute value without retaining its sign. Keep repeated cycles identifiable: repeat observations do not justify deleting an inconvenient earlier result. Record any departure from the procedure so the reviewer can assess its effect.
Do not authorise an offset adjustment from the average alone. Preserve the as-found results and ask the responsible calibration team to assess the paired errors against the manufacturer’s requirements and the application’s acceptance criteria. A useful enquiry to ProSense includes the model, range, pressure type, reference details, procedure, individual rising and falling readings, limits and intended use.
Sources and product information
Need help with your application?
ProSense Instruments - explore instruments, sensors, data loggers and test equipment.
ProSense Solutions - discuss complete monitoring and automation systems.
ProSight Software - explore cloud dashboards, alarms and reporting. Device integration must be confirmed for your application.
Instrument enquiries: sales@prosenseinstruments.com.au | 1800 560 854
Project enquiries: sales@prosense.com.au. Software enquiries: info@prosightsoftware.com.




Comments