Same Conductivity, Different TDS? Check the Conversion Factor Before Recalibrating

Nov 4
3 min read

Two handheld meters show the same conductivity, yet one reports 500 ppm TDS and the other 700 ppm. Replacing a probe or recalibrating immediately can obscure a simpler explanation: the instruments may be applying different conversion factors to the same electrical measurement.


For a useful comparison, first establish whether the disagreement is in conductivity, its temperature basis, or the calculation that produces TDS. Keep the original readings and settings. A matching display after an unexplained adjustment is not evidence that the underlying measurement was wrong.


Original conceptual illustration of two generic water-quality meters comparing conductivity and derived TDS; not a specific product

Original conceptual illustration; not a specific product, optical specification, live measurement or customer installation.


Key takeaways


  • Compare conductivity and temperature basis first.

  • Record the configured conductivity-to-TDS factor.

  • Do not recalibrate solely to force derived readings to agree.


1. Identify what the instrument is reporting


A conductivity-based estimate uses a factor multiplied by conductivity at a stated reference temperature. With EC expressed in µS/cm, this gives the conventional ppm TDS indication. Record the factor alongside the number; “ppm” alone leaves an important part of the result unspecified.


Do not confuse a TDS estimate with a direct determination of every dissolved substance. The measurement guidance TDS explanation distinguishes gravimetric measurement from an estimate based on ionic conductivity. Agreement between two conductivity-derived TDS displays therefore does not independently validate the dissolved-solids content.


2. Check the exact meter, not a familiar menu


The AZ 86032 multiparameter water quality meter is one example with a configurable TDS factor. Its manufacturer specification lists a factor range of 0.30–1.00. This capability belongs to the specified model; it should not be assumed for every pen tester or multiparameter instrument.



The AZ 86032 operating manual identifies P50 (tdF) as the TDS factor setting, with a factory default of 0.5. It separately identifies calibration review and cell-constant review functions. Reviewing a conversion factor is a different task from changing the conductivity calibration. Confirm the manual matches the supplied instrument before using its menu descriptions.


3. A hypothetical comparison


Assume both meters report 1,000 µS/cm on the same agreed temperature basis. Meter A uses 0.50: its calculated indication is 500 ppm. Meter B uses 0.70: its indication is 700 ppm. The 200 ppm difference arises entirely from the calculation in this example.


Divide each TDS indication by its configured factor and both imply 1,000 µS/cm. That arithmetic is a consistency check, not a calibration certificate or a reason to choose whichever factor produces a preferred result. Establish the required reporting method before altering a setting.


Hypothetical conductivity of 1000 microsiemens per centimetre multiplied by factors 0.50 and 0.70 gives derived TDS of 500 and 700 milligrams per litre

Original explanatory diagram. Numerical values are hypothetical where shown; follow the actual equipment documentation and approved application method.


4. Resolve the disagreement in a controlled order


  1. Save the evidence: record sample identity, time, meter and probe models, displayed EC and TDS, units, temperature and configured factor. Photograph settings where useful.

  2. Compare conductivity first. Convert units explicitly: 1 mS/cm equals 1,000 µS/cm. Confirm both results use the same temperature reference and compensation basis.

  3. Check measurement quality before comparing calculated values. Follow each probe’s instructions for immersion, rinsing and stabilisation; use suitable fresh standards when checking calibration.

  4. Compare the configured TDS factors and the method required by the application. If the factor is fixed or unavailable, record that limitation rather than assuming a default.

  5. If an authorised setting change is justified, retain before-and-after results and the reason. Recheck the sample and document the method used for subsequent records.


If EC itself disagrees, investigate those causes; changing the TDS factor cannot repair the original conductivity measurement.


5. Keep a usable comparison record


A worksheet should distinguish measured EC, its units and temperature basis, calculated TDS and its factor, and any independent laboratory result. Mark unresolved differences clearly. Avoid overwriting historical TDS values after a factor change: otherwise an apparent process trend can become a change in reporting method.


For application support, send ProSense the two meter models, paired EC/TDS readings, settings, sample type and required reporting method. Ask for help confirming a suitable instrument and documented comparison procedure. Neither a low TDS number nor this comparison alone establishes drinking-water safety, discharge compliance or suitability for a critical process.